4 Description of function
15 Pump Combinations Operation Manual, Version E
4 Description of function
This chapter explains the basic principle of the Pump Systems and provides a functional description of
the assemblies.
The design of the pump systems enables the delivery of organic as well as aqueous media without
metal contact through a pulsation-poor, 3-piston pump. The modular design of the pump systems
offers various combination options:
Pump system 1:
Pump Controller C-610 and 1 Pump Module C-601
Pump system 2:
Pump Manager C-615 and 1 Pump Module C-601 or C-605
Pump system 3:
Pump Manager C-615 and 2 Pump Modules C-601 or C-605
Pump system 4:
Control Unit C-620 and 2-4 Pump Modules C-601 or C-605
4.1 Functional principle
Pump Controller C-610:
Control unit for isocratic separations with one Pump Module C-601.
Communication with an optional Fraction Collector
C-660.
Simple entry via dial and display of delivery amount on a backlit LCD display.
Pump Manager C-615:
Control unit for one or two pump modules (C-601/605).
Communication with an optional Fraction Collector C-660.
Pressure monitoring.
Control of two optional valves for solvent selection.
Optional level monitoring.
Control Unit C-620:
Control Unit in combination with the software SepacoreControl is able to control your chromatography
system.
Pump Module C-601:
Compact, pulsation-poor, 3-piston pump to a max. of
10 bar, chemically inert and biocompatible.
Pump Module C-605:
Compact, pulsation-poor, 3-piston pump to a max. of
50 bar, chemically inert and biocompatible.