Buchi NIRFlex N-500 User Manual Page 1

Browse online or download User Manual for Accessories for air Buchi NIRFlex N-500. BUCHI NIRFlex N-500 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - Operation Manual

Operation Manual093020 enNIRFlex®N�500

Page 2

2 Safety10 NIRFlex N-500 Operation Manual, Version F2.5 Product safetyThe NIRFlex N-500 has been designed and built in accordance with state-of-the-

Page 3 - Table of contents

2 Safety11 NIRFlex N-500 Operation Manual, Version F NOTICERisk of instrument damage by mechanical shocks.Do not move the instrument when it is activ

Page 4

2 Safety12 NIRFlex N-500 Operation Manual, Version FDevice labels Meaning LocationLASER emission Sticker inside housing, at the LASER module2.5.3 Pe

Page 5

3 Technical data13 NIRFlex N-500 Operation Manual, Version F3 Technical dataThis section introduces the reader to the instrument specifications and t

Page 6 - 1 About this manual

3 Technical data14 NIRFlex N-500 Operation Manual, Version F3.2.2 Computer requirementsThe spectral data are collected in the NIRWare database on th

Page 7 - 2 Safety

3 Technical data15 NIRFlex N-500 Operation Manual, Version FTechnical data NIRFlex Solids Transmittance (cont.)Photometric linearity The “addition of

Page 8 - SIGNAL WORD

3 Technical data16 NIRFlex N-500 Operation Manual, Version F3.4 Fiber-optic measuring cells and accessories3.4.1 NIRFlex Fiber Optic SolidsTechnica

Page 9

3 Technical data17 NIRFlex N-500 Operation Manual, Version F3.4.3 NIRFlex Fiber Optic SMATechnical data NIRFlex Fiber Optic SMADetector extended ran

Page 10 - 2.5 Product safety

4 Description of function18 NIRFlex N-500 Operation Manual, Version F4 Description of functionThis section explains the basic principle of the instr

Page 11

4 Description of function19 NIRFlex N-500 Operation Manual, Version FReference laser controlThe laser deliveres a constant, wavelength-stable beam of

Page 13 - 3 Technical data

4 Description of function20 NIRFlex N-500 Operation Manual, Version FData processing and interferogram analysisThe NIR light interacts with the sampl

Page 14

4 Description of function21 NIRFlex N-500 Operation Manual, Version F4.2.1 Transflectance modeTranslucent and opaque liquids can be analyzed via tran

Page 15

4 Description of function22 NIRFlex N-500 Operation Manual, Version FNOTEWith special accessories, reflection-mode probes can be equipped for transflec

Page 16

4 Description of function23NIRFlex N-500 Operation Manual, Version F4.2.3 Transmission modeNIR light is sent through a defined pathlength of sample m

Page 17

4 Description of function24 NIRFlex N-500 Operation Manual, Version F4.3 Basic instrumentThe basic instrument consists of the spectrometer unit with

Page 18 - 4 Description of function

4 Description of function25NIRFlex N-500 Operation Manual, Version F4.3.2 Laser unitThe laser unit comprises:Laser tube•High voltage power supply•

Page 19

4 Description of function26 NIRFlex N-500 Operation Manual, Version FAbout the external PCThe external PC must run Windows® XP or Windows® 7 as opera

Page 20 - (for NIR light)

4 Description of function27 NIRFlex N-500 Operation Manual, Version F4.5 About the softwareThe NIRWare software meets industrial standards in terms

Page 21 - Detector

4 Description of function28 NIRFlex N-500 Operation Manual, Version F4.5.3 NIRCal chemometric softwareNIRCal is a modern chemometric software packag

Page 22 - NIR light

4 Description of function29 NIRFlex N-500 Operation Manual, Version F4.6.2 NIRFlex SolidsList of NIRFlex Solids add-onsSpinner / Petri dish add-onVi

Page 23

Table of contents3 NIRFlex N-500 Operation Manual, Version ETable of contents1 About this manual . . . . . . . . . . . . . . . . . .

Page 24 - 4.3 Basic instrument

4 Description of function30 NIRFlex N-500 Operation Manual, Version F1234567891011121314151617a Detectorb Magnet sensorc Status LEDd START buttone ST

Page 25 - Standard wheel

4 Description of function31 NIRFlex N-500 Operation Manual, Version F4.6.4 Vial add-on for NIRFlex SolidsNIRFlex Solids with vial add-onThe NIRFlex

Page 26 - Basic Advanced

4 Description of function32NIRFlex N-500 Operation Manual, Version F4.6.6 XL add-on for NIRFlex SolidsNIRFlex Solids with XL add-onThe NIRFlex Solid

Page 27 - 4.5 About the software

4 Description of function33 NIRFlex N-500 Operation Manual, Version F4.6.8 XL add-on for B+L sample cups for NIRFlex SolidsThe NIRFlex Solids with t

Page 28

4 Description of function34 NIRFlex N-500 Operation Manual, Version F4.6.10 Spinner add-onNIRFlex Solids in combination with the Spinner add-on is t

Page 29

4 Description of function35 NIRFlex N-500 Operation Manual, Version F4.6.12 NIRFlex Solids TransmittanceNIRFlex Solids TransmittanceIt is a major re

Page 30

4 Description of function36 NIRFlex N-500 Operation Manual, Version FTransmission measurements offer the advantage of collecting the information from

Page 31

4 Description of function37 NIRFlex N-500 Operation Manual, Version FThe picture shows the SST sample plate.The picture shows an examle of a 30-place

Page 32 - XL add-on with iris aperture

4 Description of function38 NIRFlex N-500 Operation Manual, Version F321465151413789101112a STOP buttonb START buttonc Status LEDd Detectore Optic 2f

Page 33 - External Reference

4 Description of function39 NIRFlex N-500 Operation Manual, Version F4.6.15 NIRFlex Fiber Optic SolidsNIRFlex Fiber Optic SolidsThe NIRFlex Fiber Op

Page 34 - Flow-cell adapter add-on

Table of contents4NIRFlex N-500 Operation Manual, Version E4.6.3 Petri dish add-on for NIRFlex Solids . . . . . . . . . . . . . . . .

Page 35

4 Description of function40 NIRFlex N-500 Operation Manual, Version F321910161718456111213147815a Magnet sensorsb LEDs: OK, not OKc START buttond Han

Page 36 - Sensor position and beam path

4 Description of function41 NIRFlex N-500 Operation Manual, Version F4.6.16 Transflectance adapter for NIRFlex Fiber Optic SolidsThe transflectance ad

Page 37 - NIRFlex Liquids

4 Description of function42 NIRFlex N-500 Operation Manual, Version FThe LEDs at the handle of the fiber optic probe indicate the status of the system

Page 38

4 Description of function43 NIRFlex N-500 Operation Manual, Version F4.6.18 NIRFlex Fiber Optic SMAThe NIRFlex Fiber Optic SMA is the ideal configura

Page 39 - NIRFlex Fiber Optic Solids

5 Putting into operation44 NIRFlex N-500 Operation Manual, Version F5 Putting into operationThis section describes how the instrument is installed a

Page 40

5 Putting into operation45 NIRFlex N-500 Operation Manual, Version F213In the front of the main cover of NIRFlex N-500 there is an opening which is h

Page 41 - NIRFlex Fiber Optic Liquids

5 Putting into operation46 NIRFlex N-500 Operation Manual, Version FThe instrument must be installed:indoor only with a clearance of 15 cm minimum to

Page 42

5 Putting into operation47 NIRFlex N-500 Operation Manual, Version FNOTE To cut the power in case of an emergency by unplugging, the instrument or an

Page 43 - NIRFlex Fiber Optic SMA

5 Putting into operation48 NIRFlex N-500 Operation Manual, Version FLAN setup of the PCTo establish communication, set up the IP of the connected PC

Page 44 - 5 Putting into operation

5 Putting into operation49 NIRFlex N-500 Operation Manual, Version F5.5 Installing the measuring cellsAll different measuring cells are provided wit

Page 45

Table of contents5 NIRFlex N-500 Operation Manual, Version E7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . .

Page 46

5 Putting into operation50 NIRFlex N-500 Operation Manual, Version F5.6 Installing the cover of the NIRFlex Solids TransmittancebaInsert the tilted

Page 47

5 Putting into operation51 NIRFlex N-500 Operation Manual, Version FAdditional handling instructionsInstall the measuring cell with the top cover •c

Page 48

5 Putting into operation52 NIRFlex N-500 Operation Manual, Version FIf this is not the case, proceed as follows: Slide the adjustment ring over the t

Page 49 - Installing a measuring cell

5 Putting into operation53 NIRFlex N-500 Operation Manual, Version F Adjusting the pathlength Now turn the adjustment ring upwards until there is no

Page 50

5 Putting into operation54 NIRFlex N-500 Operation Manual, Version F5.8 Installing measurement equipment to the NIRFlex Fiber Optic SMATo install th

Page 51

5 Putting into operation55 NIRFlex N-500 Operation Manual, Version F5.9 Starting up the systemAn automatic initialization process starts as soon as

Page 52

5 Putting into operation56 NIRFlex N-500 Operation Manual, Version F5.9.4 Signal-to-noise ratio determinationWith this test the ratio of signal powe

Page 53 - Exchanging the spacer ring

5 Putting into operation57 NIRFlex N-500 Operation Manual, Version FInternal and External Reference measurements provided by the measuring cells/add-

Page 54

6 Operation58 NIRFlex N-500 Operation Manual, Version F6 OperationThis section describes how to start a measurement and how to prepare a reference m

Page 55 - 5.9 Starting up the system

6 Operation59 NIRFlex N-500 Operation Manual, Version F6.2 General recommendations for measuring liquidsFor sample measurements with the NIRFlex Liq

Page 56 - 5.10 Reference measurements

1 About this manual6 NIRFlex N-500 Operation Manual, Version F1 About this manualThis manual describes the NIRFlex N-500 including its standard soft

Page 57

6 Operation60 NIRFlex N-500 Operation Manual, Version F6.4 NIRFlex Solids6.4.1 Reference measurement for the Petri dish add-onA pop-up will remind

Page 58 - 6 Operation

6 Operation61 NIRFlex N-500 Operation Manual, Version FTypical samples for the NIRFlex Solids Transmittance include tablets, hard gelatin capsules, p

Page 59 - 6.3 Starting a measurement

6 Operation62 NIRFlex N-500 Operation Manual, Version F! CAUTIONMinor or medium burns by hot surfacesDo not touch the cell chambers•NOTEThe software

Page 60 - 6.4 NIRFlex Solids

6 Operation63 NIRFlex N-500 Operation Manual, Version F! WARNINGRisk of fire by highly flammable materialsDo not insert highly flammable materials or sa

Page 61 - 6.6 NIRFlex Liquids

6 Operation64 NIRFlex N-500 Operation Manual, Version F6.7.2 NIRFlex Fiber Optic LiquidsFor most of the applications a pathlength of 1 or 2 mm is th

Page 62

7 Maintenance and repairs65 NIRFlex N-500 Operation Manual, Version F7 MaintenanceThis section gives instructions on all maintenance work to be perf

Page 63 - 6.7 NIRFlex Fiber Optics

7 Maintenance and repairs66 NIRFlex N-500 Operation Manual, Version FIn order to avoid contamination by sample material between the measurements, cle

Page 64

7 Maintenance and repairs67 NIRFlex N-500 Operation Manual, Version F NOTICERisk of damage at the External ReferenceDo not drop the External Referenc

Page 65 - 7 Maintenance

7 Maintenance and repairs68 NIRFlex N-500 Operation Manual, Version FNOTE Exact cleaning procedures tailored to the indiviudal sample composition hav

Page 66

7 Maintenance and repairs69 NIRFlex N-500 Operation Manual, Version FMyNadiaFilesSend this file to your customer service organisation for evaluation.

Page 67

2 Safety7 NIRFlex N-500 Operation Manual, Version F2 SafetyThis section introduces the safety concept of the instrument and contains general rules o

Page 68 - Test result Nadia

7 Maintenance and repairs70 NIRFlex N-500 Operation Manual, Version F! CAUTIONRisk of minor burns by hot lamp module.Switch off the device first and u

Page 69 - MyNadiaFiles

7 Maintenance and repairs71 NIRFlex N-500 Operation Manual, Version F NOTICERisk of instrument damage by ingress of particles.Check proper seating of

Page 70 - Replacing the lamp module

7 Maintenance and repairs72 NIRFlex N-500 Operation Manual, Version FWorkflow to reset the lifetime counter of the primary lamp moduleTo establish the

Page 71 - Assembly setup dialog

7 Maintenance and repairs73 NIRFlex N-500 Operation Manual, Version F7.5 Laser unit replacementThe LASER unit needs to be replaced at failure. This

Page 72

7 Maintenance and repairs74 NIRFlex N-500 Operation Manual, Version F Disconnect the electric lasers power supply •plug. Loosen the two screws which

Page 73 - 7.5 Laser unit replacement

7 Maintenance and repairs75 NIRFlex N-500 Operation Manual, Version FResetting the laser lifetime counterAfter LASER unit replacement the following m

Page 74 - Removing the LASER unit

7 Maintenance and repairs76 NIRFlex N-500 Operation Manual, Version FThe following screen indicates the laser lifetime counter is set to zero:Hours o

Page 75

7 Maintenance and repairs77 NIRFlex N-500 Operation Manual, Version F7.7 Main fuse replacementThe primary fuses are located at the rear of the instr

Page 76 - Replacing the filter pads

7 Maintenance and repairs78 NIRFlex N-500 Operation Manual, Version F7.8 NIRFlex SolidsIt is possible that solids or pasty samples are spilled into

Page 77 - Replacing the primary fuses

7 Maintenance and repairs79 NIRFlex N-500 Operation Manual, Version F7.10 Customer serviceOnly authorised service personnel is allowed to perform re

Page 78 - 7.9 NIRFlex Liquids

2 Safety8 NIRFlex N-500 Operation Manual, Version F2.4 Safety warnings and safety signs used in this manualDANGER, WARNING, CAUTION and NOTICE are s

Page 79 - 7.10 Customer service

8 Storage, transport and disposal80 NIRFlex N-500 Operation Manual, Version F8 Storage, transport and disposalThis section instructs how to handle t

Page 80 - 8.2 Disposal

8 Storage, transport and disposal81 NIRFlex N-500 Operation Manual, Version F Health and Safety Clearance Health and Safety Clearance_20081110.doc_2

Page 81 - Health and Safety Clearance

9 Spare parts82 NIRFlex N-500 Operation Manual, Version F9 Spare parts and accessoriesThis section lists spare parts, accessories and options of the

Page 82 - 9.1 Scope of delivery

9 Spare parts83 NIRFlex N-500 Operation Manual, Version F9.1.2 Measuring cellsNIRFlex SolidsProduct Order numberFor measurements in diffuse reflectio

Page 83 - Product Order number

9 Spare parts84 NIRFlex N-500 Operation Manual, Version FNIRFlex Fiber Optic SolidsProduct Order numberFor measurements in diffuse reflection2 m fiber

Page 84 - N I R War e + NI R Cal

9 Spare parts85 NIRFlex N-500 Operation Manual, Version F9.1.4 Standard accessoriesExampleStandard accessoriesProduct Order number1 Power supply cab

Page 85 - 93124

9 Spare parts86 NIRFlex N-500 Operation Manual, Version F9.1.5 Optional accessoriesOptional accessoriesProduct Order numberSet of 2 quartz glass cuv

Page 86

9 Spare parts87 NIRFlex N-500 Operation Manual, Version FOptional accessories (cont.)Product Order numberTransflectance cover or high performance cup

Page 87

9 Spare parts88 NIRFlex N-500 Operation Manual, Version FOptional accessories (cont.)Product Order numberHolder for fiber optic probe 46273Barcode rea

Page 88

9 Spare parts89 NIRFlex N-500 Operation Manual, Version FOptional accessories (cont.)Product Order numberStamp for Hellma cuvettes 11056459Flow add-o

Page 89 - 9.2 NIRFlex spectrometer

2 Safety9 NIRFlex N-500 Operation Manual, Version FSymbol MeaningExplosive gases, explosive environmentHot item, hot surfaceDevice damageFragile comp

Page 90 - 9.3 NIRFlex Solids

9 Spare parts90 NIRFlex N-500 Operation Manual, Version F9.3 NIRFlex SolidsSpare parts NIRFlex SolidsProduct Order numberFuse 2.0 AT 02998External R

Page 91 - 9.5 NIRFlex Liquids

9 Spare parts91 NIRFlex N-500 Operation Manual, Version F9.4 NIRFlex Solids TransmittanceSpare parts NIRFlex Solids TransmittanceProduct Order numbe

Page 92 - 9.9 Transflectance adapter

9 Spare parts92 NIRFlex N-500 Operation Manual, Version F9.8 NIRFlex Fiber Optic SMASpare parts Fiber Optic SolidsProduct Order numberFuse 2.0 AT 02

Page 93

10 Declarations and requirements93 NIRFlex N-500 Operation Manual, Version F10 Declarations and requirements10.1 FCC requirements (for USA and Cana

Page 94

10 Declarations and requirements94 NIRFlex N-500 Operation Manual, Version F10.2 Declaration of conformity

Page 96 - Flawil 1 / Switzerland

BÜCHI Labortechnik AGCH-9230 Flawil 1 / SwitzerlandT +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com Quality in your hands

Comments to this Manuals