4.2.1 Transflectance mode
Translucent and opaque liquids can be analyzed via transflectance mode. The light penetrates the
liquid, is diffusely reflected by the reference plate and passes the sample a second time. The trans-
flected rays contain the spectral information of the sample.
Factors that can influence the measurement:
Not enough sample material available•
Air bubbles in the measurement path and under the transflectance cover•
Sample not homogeneous or representative•
Temperature of sample material•
Sample cup or transflectance cover inadequate (e.g. cup material and thickness, blind spots, dirt •
etc.)
NIRFlex Solids:
Diffuse reflecting / reference plate
Small, defined gap for liquid samples
Detector
NIR light
Sample cup / petri dish
NIRFlex Fiber Optic Solids / SMA:
Diffuse reflecting / reference plate
Small gap for liquid samples
Fiber optics (example)