4 Description of function
41 NIRFlex N-500 Operation Manual, Version F
4.6.16 Transflectance adapter for NIRFlex Fiber Optic Solids
The transflectance adapter is an optional accessory for the NIRFlex Fiber Optic Solids and enables to
measure liquid samples, pastes and transparent films.
The transflectance measurement principle is a combination of transmission and diffuse reflection: The
light coming from the optical fibers penetrates the liquid sample in the gap between the probe window
and the transflectance adapter window (½ pathlength), is diffusely reflected at the white standard layer
behind the adapter window, passes the sample liquid in the gap a second time and goes back into
the optical fibers.
The transflectance adapter is suitable for the analysis of viscous, non-agressive liquids. Thanks to the
reproducible, adjustable pathlength it is suitable both, for the meausurement of slightly and of strongly
absorbing samples.
The transflectance technique is a good compromise for customers, who only analyze liquids occasio-
nally.
As the measurements are carried out in the original sample containers, the probe head needs to be
dipped into the sample.
Transflectance adapter
4.6.17 NIRFlex Fiber Optic Liquids
NIRFlex Fiber Optic Liquids
The NIRFlex Fiber Optic Liquids is the ideal configuration to measure liquid samples in their original
container.
The transmission probe operates with monofibers.
As the measurements are carried out in the original sample containers, the probe head needs to be
dipped into the samples.
The handle is equipped with a remote control for initiating the measurements remotely and displaying
the results using LEDs on the handle.