u Installtheloop(2)betweentheportno2and5.
HAF560Injectionloop,2mL
5.2.3 TubingFlash/prepcolumn,installation
Tasks
u Installluerlockadapters(6)and(4).
HAF450Adapterluerlockmaleto1/4"–28male
HAF630Adapter1/4"–28femaletoluermale
u Installthetube(1)betweentheportno3ofthe6wayelectricalvalveand
columninlet.
HAF830TubingETFE,1.6mmID
u Installthetube(3)betweenthecolumnoutletandinletoftheUVDetector
module.Thelongnutofthetubemustbeconnectedwiththecolumn.
HAF860Tubing,ETFE,1.6mmID,plasticconnectors
u Installthetube(2)betweenthefractioncollectorinlettotheUVDetector
moduleoutlet.
NOTE
Thetubeisalreadypreinstalledontheinletintherearofthefractioncollector
module.
u Installthetube(5)betweentheportno6ofthe6wayelectricalvalveand
wastebottle.
HAF480TubingETFE,valveport6towaste,100cm,1.6mmID
u Insertthewastetubefromtherearofthefractioncollectorintothewaste
bottle.