Buchi PrepChrom C-700 User Manual Page 95

  • Download
  • Add to my manuals
  • Print
  • Page
    / 110
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 94
BÜCHILabortechnikAG Maintenance|7
PrepChromC700 95/110
u Airpurgethesystem.
6.9 Air purge [} page85]
Tasks
u Opentheservicedoorofthepumpmodule.
u Removebothfittings(1)ontheoutletofthepump.
Thetubescannowbeliftedoffthevalveinserts(2).
Wrench3/8"
u Disconnectbothtubes(3)ontheinletofthepump.
u Removethecheckvalveinserts(2).
Wrench17mm
u Removethecheckvalvefromthecheckvalveinsert.
7.3.5.2 Checkvalve,installation
u Placethecheckvalve(2)inthecheckvalveinsert(1).
u Makesurethecheckvalvearrowshowsalwaystowardsthesolventflow
direction.
NOTE
Makesurethecheckvalveiscorrectlyinstalled,otherwisethepumpwillnot
workproperly.