u Holdtheflowcell(2)withonehandwhileliftingthelever(1)furtherup.
u Removetheflowcell.
u Flushthecellwithpuresolventandrinseextensively.
u ReassembletheUVDetectormoduleinreverseordertodisassembly.
Followontasks
u Purgethesystemandverifythereisnobaselinedrift.
7.2.3 Checkvalve,cleaning
Preliminarytasks
u Purgethesystemwithasuitablesolvent(isopropanol).
u Airpurgethesystem.
6.9 Air purge [} page85]
u 7.3.5.1 Check valve, removal [} page94]
Tasks
NOTE
Alwayscleanall4checkvalves.
u Cleanthecheckvalvesinaultrasonicbathwithmethanoloracetonefor
5minutes.
Followontasks
u 7.3.5.2 Check valve, installation [} page95]