4 Description of function
31 R-220 SE Operation Manual, Version F
d
The optional cooling water valve interrupts the water flow in
order to effectively reduce the water consumption. It must be
installed in serial after the cooling water source.
e
See master drawing for appearance
and mounting position.
The optional vacuum valve must be installed in case a 3rd
party vacuum pump is used. It is a safety device to control the
vacuum level.
f
Fuse triggered
Fuse ok
Resettable fuses. Gently push them back into prestressed posi-
tion when fuses are activated.
g
The optional foam detector is located before the expansion
vessel. In case of foam detection its signal opens the aeration
valve for an instant. This aeration pulse causes the foam to
collapse. For installation a special distribution piece is needed!
h
Alarm output with MiniDIN 4-pole socket. It supplies a “normally
open” output. See technical data and section 3.2 for more
information.
i
Mains input cable to supply the Rotavapor. Output socket to
supply a vacuum pump. Mind the voltage and maximal output
current, when installing a vacuum pump! The combination
of Rotavapor and vacuum pump must not violate the local
electrical safety regulations. Contact an electrician before
connecting any device.
j
USB service interface, socket type B.