5 Putting into operation
34 R-220 SE Operation Manual, Version F
NOTE
Secure the instrument against tilting and unintentional moving in earthquake prone regions.
5.2 Electrical connections
The Rotavapor is aimed to be installed stationary and is not equipped with a power plug. The elec-
trical installation must be performed by an electrician or equivalent specialist in any case. After the
installation procedure an electrical safety test must be performed to verify a safe system condition
such as sufficient grounding.
Mobile / stationary installation
In mobile mode the instrument is connected to mains via a power plug. The plug must be freely
accessible at any time to be unplugged in case of emergency.
In case the instrument is installed stationary (without power plug) an emergency power-off device
must be installed in the vicinity of the Rotavapor with free access granted.
Demands on the mains circuit
The mains circuit must
provide the voltage that is given on the type plate of the instrument.
be able to handle the load of the connected instruments.
be equipped with adequat fusage and electrical safety measures, in particular proper grounding.
For details about the installation requirements, see "Installation Instruction R-220SE" document.
See also technical data (section 3.2) of all components regarding the different minimum system
requirements!
NOTE
Additional electrical safety measures such as residual current breakers may be necessary to
meet local laws and regulations!
External power switches (e.g. emergency stop switches) must meet IEC 60947-1 and IEC
60947-3 requirements. Such devices must be clearly labeled and accessible at any time.
External connections and extension lines must be provided with a grounded conductor lead
(3-pole couplings, cord or plug equipment). All used power cords must meet the input power
requirements.
!
WARNING
Death or serious burns by electric current at installation.
The instrument must be installed by an electrician or equivalent specialist
After installation, the instrument must be tested for electrical safety
NOTICE
Risk of instrument damage by wrong mains supply.
External mains supply must meet the voltage given on the type plate
Check for sufficient grounding