5 Putting into operation
42 R-220 SE Operation Manual, Version F
5.5 Setting up the support rod(s)
a
b
c
d
a
Install the retaining bracket at the housing. Fasten the Torx screw well handtight.
b
Premount the headless screw with a few turns at the side thread of the bracket.
c
Remove the blind plug from the blind hole at the foot of the Rotavapor. Insert the short rod from above
through the retaining bracket into the blind hole.
d
Tighten the headless screw at the retaining bracket to safely fixate the rod. Now, the receiving flask
holder(s) can be installed.