5 Putting into operation
37 R-220 SE Operation Manual, Version F
5.3.2 Cooling water flow sensor
The switching threshold of the flow sensor can be adjusted by shifting the sensor element up and
down.
CW Flow
b
a
c
Adjustment instructions
Make sure the flow sensor is installed in an upright position! Loosen the locking screw a.
Switch on the Rotavapor and shift the sensor far downwards. The display must not show a flow
sensor error in this position.
Subsequently shift the sensor upwards until it generates a flow sensor error on the display
(E12, "FLOW"). To avoid unintentional switching, lift the sensor 2-3 more millimeters and tighten the
locking screw to fix the sensor position.
Turn on the cooling water. The error should now disappear.