8 Troubleshooting
81 R-220 SE Operation Manual, Version F
Error Additional error indicator on display Possible cause Remedy
E03 Drive unit icon flashes. The rotary drive does not reach the
set sppec value.
Drive defective or blocked.
Press the “Aerate” button to confirm
the error message.
Switch off the unit and check if
drive is blocked, rotate evaporating
flask manually.
Call service if problem is persistent.
E04 n.a. Battery voltage too low.
PB battery weak or defective.
Press the “Aerate” button to confirm
the error message.
Switch on the system over night to
allow recharging of the battery.
Call service if problem is persistent.
E05 n.a. Hardware malfunction. Restart the unit.
Call service if problem is persistent.
E06 n.a. Shortcut at aeration valve. Unplug valve and restart system for
causal research.
Exchange valve if defective or run
evaporation with manual aeration.
Call service if problem is persistent.
E07 n.a. Shortcut at cooling water valve. Unplug valve and restart system for
causal research.
Exchange valve if defective or run
evaporation without cooling water
valve.
Call service if problem is persistent.
E08 n.a. Shortcut at vacuum valve. Unplug valve and restart system for
causal research.
Exchange valve if defective.
Call service if problem is persistent.
E11 ‘SHIELD OPEN’ message
appears.
Shield not closed at the time of
flask rotation start.
Close bath shield.
Press the “Aerate” button to confirm
the error message.
Call service if problem is persistent.