5 Putting into operation
36 R-220 SE Operation Manual, Version F
5.3.1 Level sensors
Description of function:
The capacitive level sensors are recommended for setups with two receiving flasks to achieve unin-
terruptible evaporation over a longer period and for larger product volumes. The sensitivity of the
capacitive level sensors must be adjusted to safely detect the liquid level between different products.
Installation instructions:
Mount the sensor at the vessel and connect it at the rear-side connection panel a.
Fill in some solvent.
Shift the sensor until it is located below the liquid level.
The red LED at the sensor should now be off.
If the sensor does not safely detect the liquid:
Use a small screwdriver to set the sensitivity (with the small adjustment screw) b of the
sensor.
NOTE
The sensor detects a liquid when the red LED is off. The message "Level 1" or "Level 2" is
displayed on the Rotavapor display, depending on the tested sensor loop.
LEVEL 2
LEVEL 1
Front view
Rear view The crosshair marks the sensitive area
Status LED
Adjustment screw
b