5 Putting into operation
33 R-220 SE Operation Manual, Version F
5 Putting into operation
This chapter describes how the instrument has to be installed. It also gives instructions for the initial
startup.
NOTE
Inspect the instrument for damages during unpacking. If necessary, prepare a status report imme-
diately to inform the postal company, railway company or transportation company. Keep the original
packaging for future transportation.
5.1 Installation site
Put the instrument on a stable, horizontal surface. Consider the maximum product dimensions and
weight. Obtain the environmental conditions as described in section 3.2, technical data. Installation
on the dedicated mobile trolley (order no. 041257) or the BUCHI chiller model F-125 is also possible.
Installation prerequisites:
• Do not place any objects on top or below the instrument or parts of it.
• The instrument must be installed with 10 -15 cm clearance to any other objects or walls to allow
sufficient cooling.
• Do not store containers, chemicals or other items behind the instrument.
• When working with hazardous solvents or media, install the instrument inside a fume hood.
!
WARNING
Death or serious injuries by use in explosive environments.
• Do not operate the instrument in explosive environments
• Do not operate the instrument with explosive gas mixtures
• Before operation, check all gas connections for correct installation
• Directly withdraw released gases and gaseous substances by sufficient ventilation
!
CAUTION
Risk of minor or moderate injury by heavy weight of the instrument.
• Consult a second person to transport the instrument
• Do not drop the instrument
• Place the instrument on a stable, even and vibration-free surface
• Keep limbs out of crushing zone
NOTICE
Risk of instrument damage by liquids or mechanical shocks.
• Do not spill liquids over the instrument or its components
• Do not move the instrument when it is loaded with sample liquid
• Do not drop the instrument or its components
• Keep external vibrations away from the instrument
• Safely attach the instrument to the bench in earthquake prone regions
• Do not operate the instrument without the protection cover installed at the front