Test results
Contaminant Concentration germs/ml Sample 1 germs/ml Sample 2 germs/ml Sample 3 germs/ml
Escherichia coli ATCC 25922 10
3
0 0 0
Escherichia coli ATCC 25922 10
4
0 0 0
Escherichia coli ATCC 25922 10
5
0 0 0
Escherichia coli ATCC 25922 10
6
0 0 0
Staphylococcus aureus ATCC 25923 10
3
0 0 0
Staphylococcus aureus ATCC 25923 10
4
0 0 0
Staphylococcus aureus ATCC 25923 10
5
0 0 0
Staphylococcus aureus ATCC 25923 10
6
0 0 0
Saccharomyces cerevisiae NCYC 79 10
3
0 0 0
Saccharomyces cerevisiae NCYC 79 10
4
0 0 0
Saccharomyces cerevisiae NCYC 79 10
5
0 0 0
Saccharomyces cerevisiae NCYC 79 10
6
0 0 0
Geobac. stearothermophilus ATCC 7953 10
3
40 0 0
Geobac. stearothermophilus ATCC 7953 10
4
560 - -
Geobac. stearothermophilus ATCC 7953 10
5
3600 0 0
Geobac. stearothermophilus ATCC 7953 10
6
46000 0 0
5.12 General installation instructions for glassware
Small tolerances of glassware parts are conditional of manufacturing. Thus, it is necessary to indi-
vidually adjust fixtures and holders at installation to avoid stress at the components.
It is recommended to install a glassware setup starting with the distribution piece at the gearbox. All
other parts will subsequently be aligned towards the Rotavapor via the distribution piece. The drawing
below shows an example installation of a complex setup. For other setups proceed adaptive (e.g.
leave parts / steps out).